With 6 mM Tartaric asid and 4 mM oxalic asid aqueous solution as the eluent, transision metals, such as nickel, zinc, and cobalt can be analyzed using IC YS-50.
Sample : 20 μL
1. Zn2+ 10 mg/L
2. Co2+ 10 mg/L
3. Fe2+ 10 mg/L
4. Mn2+ 10 mg/L
5. Cd2+ 10 mg/L
6. Pb2+ 30 mg/L
Column : Shodex IC YS-50 (4.6 mm I.D. x 125 mm) Eluent : 6 mM Tartaric acid + 4 mM Oxalic acid aq. Flow rate : 1.0 mL/min Detector : Non-suppressed conductivity Column temp. : 40 °C
Sample Name Index
Operation Manual / Certificate of Analysis