Please read the column cleaning method for your column.
RSpak DS-613, RSpak DS-413, RSpak DE-613, RSpak DE-413 and RSpak DE-213
For an efficient cleaning, reverse the flow direction.
| Product name | DS-613 | DS-413 | DE-613 | DE-413 | DE-213 |
|---|---|---|---|---|---|
| Flow rate | ≤ 0.5 mL/min | ≤ 0.3 mL/min | ≤ 0.5 mL/min | ≤ 0.5 mL/min | ≤ 0.1 mL/min |
Method 1: Basic solvent wash (Apply after using the column under acidic to neutral condition for an extended period of time.)
Introduce 10-mM aqueous sodium hydroxide solution/acetonitrile = 80/20. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
Method 2: Acid solvent wash (Apply after using the column under basic to neutral condition for an extended period of time.)
Introduce 50-mM aqueous phosphoric acid solution/acetonitrile = 90/10. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
Method 3: Organic solvent wash (Apply after using the column under highly aqueous condition for an extended period of time.)
Introduce water/acetonitrile = 20/80. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
- (Attention)
Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
RSpak DM-614
For an efficient cleaning, reverse the flow direction.
Set the flow rate at 0.3 mL/min, and introduce 50 mL of methanol, acetonitrile or water. (Cleaning time: 165 minutes)
- (Attention)
- Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
RSpak NN-814
Method: adsorbing ionic componentsIntroduce the eluent and inject 20 - 50 μL of 2-M nitric acid aqueous solution several times.
RSpak JJ-50 2D
For an efficient cleaning, reverse the flow direction.
| Washing solution (Flow rate: 0.1 mL/min) | Cleaning time (min) | |
|---|---|---|
|
1 |
H2O | 6 |
| 2 | 0.1 M NaOH aq. solution | 75 |
| 3 | H2O | 12 |
| 4 | Eluent | 40 |
| Washing solution (Flow rate: 0.1 mL/min) | Cleaning time (min) | |
|---|---|---|
| 1 | H2O | 6 |
| 2 | 0.1 M HNO3 aq. solution | 75 |
| 3 | H2O | 6 |
| 4 | 0.1 M NaOH aq. solution | 75 |
| 5 | H2O | 12 |
| 6 | Eluent | 40 |
| Washing solution (Flow rate: 0.1 mL/min) | Cleaning time (min) | |
|---|---|---|
| 1 | H2O | 25 |
| 2 | H2O/CH3CN = 95/5 | 20 |
| 3 | CH3CN | 100 |
| 4 | H2O | 50 |
| 5 | Eluent | 40 |
- (Attention)
- Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
- Strong alkaline solvents such as 0.1-M aqueous sodium hydroxide solution can damage the detector cell. Do not connect the detector while regenerating the column and collect elute directly from the column outlet to the waste container.