Please read the column cleaning method for your column.
RSpak DS-613, RSpak DS-413, RSpak DE-613, RSpak DE-413 and RSpak DE-213
For an efficient cleaning, reverse the flow direction.
Product name | Flow rate | Product name | Flow rate | Product name | Flow rate |
---|---|---|---|---|---|
RSpak DS-613 | ≤ 0.5 mL/min | RSpak DS-413 | ≤ 0.3 mL/min | RSpak DE-613 | ≤ 0.5 mL/min |
RSpak DE-413 | ≤ 0.5 mL/min | RSpak DE-213 | ≤ 0.1 mL/min |
Introduce 10-mM aqueous sodium hydroxide solution/acetonitrile = 80/20. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
Introduce 50-mM aqueous phosphoric acid solution/acetonitrile = 90/10. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
Introduce water/acetonitrile = 20/80. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
- (Attention)
Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
For an efficient cleaning, reverse the flow direction.
Set the flow rate at 0.3 mL/min, and introduce 50 mL of methanol, acetonitrile or water. (Cleaning time: 165 minutes)
- (Attention)
- Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
Introduce the eluent and inject 20 - 50 μL of 2-M nitric acid aqueous solution several times.
For an efficient cleaning, reverse the flow direction.
Washing solution (Flow rate: 0.1 mL/min) | Cleaning time (min) | |
---|---|---|
1 |
H2O | 6 |
2 | 0.1 M NaOH aq. solution | 75 |
3 | H2O | 12 |
4 | Eluent | 40 |
Washing solution (Flow rate: 0.1 mL/min) | Cleaning time (min) | |
---|---|---|
1 | H2O | 6 |
2 | 0.1 M HNO3 aq. solution | 75 |
3 | H2O | 6 |
4 | 0.1 M NaOH aq. solution | 75 |
5 | H2O | 12 |
6 | Eluent | 40 |
Washing solution (Flow rate: 0.1 mL/min) | Cleaning time (min) | |
---|---|---|
1 | H2O | 25 |
2 | H2O/CH3CN = 95/5 | 20 |
3 | CH3CN | 100 |
4 | H2O | 50 |
5 | Eluent | 40 |
- (Attention)
- Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
- Strong alkaline solvents such as 0.1-M aqueous sodium hydroxide solution can damage the detector cell. Do not connect the detector while regenerating the column and collect elute directly from the column outlet to the waste container.