Glyphosate and glufosinate which are phosphorus-containing amino acid type herbicides were analyzed with IC/ICP-MS using IC SI-90 4E (a column for anion chromatography).
Sample : 20 μL
1. Glufosinate
2. Glyphosate
Column : Shodex IC SI-90 4E (4.0 mm I.D. x 250 mm) Eluent : (A) 0.3 mM NaOH aq. (B) 0.3 mM NaOH + 20 mM Na2CO3 aq. Gradient: 0 to 20 min, A/B=95/5 20 to 30 min, A/B=95/5 to 50/50 30 to 40 min, A/B=50/50 Flow rate : 1.0 mL/min Suppresor : Dionex AMMS-ICS II Detector : ICP/MS, P at m/z=31 Column temp. : Room temp.
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- LC/MS/MS Analysis of Glyphosate and Glufosinate (VT-50 2D)
- LC/MS Analysis of Glyphosate and Glufosinate (VT-50 2D)
- Analysis of Glyphosate in Serum
- Ethoxyquin (ODP-50 4E)
- Insecticides (Neonicotinoids) Analysis (DE-413)
- Paraquat and Diquat (ES-502C 7C)
- Analysis of Agricultural Chemicals by LC/MS (DE-213)
- Ultra-Rapid Analysis of Commonly Used Pesticides on Golf Courses (C18U 2B)