Scroll
GF-HQ
  • Polymer-based SEC columns with high solvent durability
  • Works well with both aqueous and organic solvents
(Standard columns) [Housing material: SUS]Scroll
Product code Product name Plate number
(TPN/column)
Particle size (μm) Column size (mm)
I.D. x Length
F7600000

Asahipak GF-210 HQ

≥ 19,000 5 7.5 x 300
F7600001 Asahipak GF-310 HQ ≥ 19,000 5 7.5 x 300
F7600002 Asahipak GF-510 HQ ≥ 19,000 5 7.5 x 300
F7600004 Asahipak GF-7M HQ ≥ 13,000 9 7.5 x 300
F6710018 Asahipak GF-1G 7B (guard column) 9 7.5 x   50
F7600110 MSpak GF-310 4D ≥ 10,000 5 4.6 x 150
GF-7M HQ is a mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution.
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
Asahipak GF-210 HQ 9.0 0.4 - 0.6 1.0 4 - 60 2 - 9 H2O
Asahipak GF-310 HQ 7.0 0.4 - 0.6 1.0 4 - 60 2 - 9 H2O/CH3OH=70/30
Asahipak GF-510 HQ 6.5 0.4 - 0.6 1.0 4 - 60 2 - 9 H2O/CH3OH=70/30
Asahipak GF-7M HQ 4.5 0.4 - 0.6 1.0 4 - 60 2 - 9 H2O/CH3OH=70/30
Asahipak GF-1G 7B - - - 4 - 60 2 - 9 H2O/CH3OH=70/30
MSpak GF-310 4D 4.0 0.2 0.5 4 - 60 2 - 9 H2O
(Preparative columns) [Housing material: SUS]   [Preparative columns are made to order.]Scroll
Product code Product name Plate number
(TPN/column)
Particle size
(μm)
Column size (mm)
I.D. x Length
Shipping solvent Standard columns
F6810038 Asahipak GS-310 20G ≥ 14,000 13 20.0 x 500 H2O/CH3OH=70/30 GF-310 HQ
F6810039 Asahipak GS-510 20G ≥ 14,000 13 20.0 x 500 H2O/CH3OH=70/30 GF-510 HQ
F6710020 Asahipak GS-10G 7B (guard column) 20  7.5  x   50 H2O/CH3OH=70/30 (guard column)
Usable solventsScroll
Solvents Product name
GF-210 HQ GF-310 HQ GF-510 HQ GF-7M HQ
Water (0 - 0.5 M salt concentration)
Methanol
Ethanol
Acetonitrile
Tetrahydrofuran (THF)
Acetone
2-Propanol (IPA) -
Ethyl acetate
N,N-Dimethylformamide (DMF)
Dioxane -
Dimethyl sulfoxide (DMSO)
(0 -50 %)

(0 -50 %)

(0 -50 %)
N,N-Dimethylacetamide (DMAc) - - -
Chloroform
: Can be used
Target molecular weight range and exclusion limitPlease use the following tables for reference purposes only when selecting columns. Scroll
Product name Measured with pullulan (eluent: ultrapure water)  Measured with **PEG/PEO (eluent: DMF)
Target molecular weight range Exclusion limit Target molecular weight range
Asahipak GF-210 HQ 300 - 4,000           9,000 100 - 2,000
Asahipak GF-310 HQ 300 - 30,000         40,000 200 - 4,000
Asahipak GF-510 HQ 5,000 - 200,000       300,000 2,000 - 200,000
Asahipak GF-7M HQ 300 - *(10,000,000) *(10,000,000) 200 - *(10,000,000)
*( ) Estimated value
** Polyethylene glycol/Polyethylene oxide

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Related Products

Product Information (Related Information)

General Information on Packed Columns