To realize the best performance of RSpak DE-413 column, we recommend to use the following HPLC system with the column.
Device | Recommended specifications and/or setting |
---|---|
Pump | Flow rate: max. 1 mL/min. |
Injector | Low dead volume type Sample loop: max. 50 μL (5 μL loop is recommended) |
From injector to column | 0.1 to 0.13 mm I.D., 100 mm Length piping |
From column to detector | 0.1 mm I.D. piping |
Detector | Detector with a micro cell should be used. Response should be set to "FAST". |
Sample : (A); 1.5 μL, (B); 5 μL
1. Methyl ethyl ketone, 2-Butanone 0.2 %
2. Diethyl ketone, 3-Pentanone 0.2 %
3. Di-n-propyl ketone, 4-Heptanone 0.4 %
4. Di-n-butyl ketone, 5-Nonanone 0.6 %
Column : Shodex RSpak DE-413 (4.6 mm I.D. x 150 mm) Eluent : CH3CN/H2O=50/50 Flow rate : 1.0 mL/min Detector : UV (262 nm) Column temp. : 40 °C