Please read the column cleaning method for your column.
For an efficient cleaning, reverse the flow direction.
Set the flow rate at 0.5 mL/min. Introduce the eluent and inject 1 to 2 mL of 30 % (v/v) aqueous acetic acid solution several times.
For an efficient cleaning, reverse the flow direction.
Introduce an eluent with higher salt concentration (about 1 M) or 0.1 % nonionic surfactants. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
Introduce a mixture of solvent containing 50-mM eluent (buffer or aqueous salt solution) and 10 to 20 % (v/v) acetonitrile. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
- (Attention)
- Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
For an efficient cleaning, reverse the flow direction.
Introduce an eluent with higher salt concentration or 0.1-M aqueous acetic acid solution. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
Introduce a mixture of solvent containing 50-mM eluent (buffer or aqueous salt solution) and 50 % (v/v) acetonitrile or methanol. Recommended solvent volume to introduce is 5 to 10 times of the column volume.
- (Attention)
- Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.