Please read the column cleaning method for your column.
For an efficient cleaning, reverse the flow direction.
Washing solution (Flow rate: 0.5 mL/min) | Cleaning time (min) | |
---|---|---|
1 | H2O | 10 |
2 | 4 mM Methanesulfonic acid aq. solution/CH3CN = 70/30 | 40 |
3 | H2O | 10 |
4 | Eluent | 40 |
Washing solution (Flow rate: 0.5 mL/min) | Cleaning time (min) | |
---|---|---|
1 | H2O | 10 |
2 | 20 mM Methanesulfonic acid aq. solution | 40 |
3 | H2O | 20 |
4 | Eluent | 40 |
- (Attention)
- Keeping the Washing solution in the column for a long time will lead to column deterioration. Please replace the Washing solution with the eluent immediately after cleaning.
For an efficient cleaning, reverse the flow direction.
Set the flow rate at 0.3 mL/min. Introduce 50 mL of a washing solution prepared of 5-mM aqueous tartaric acid solution/acetonitrile = 50/50. (Cleaning time: 165 minutes)
Set the flow rate at 0.3 mL/min. Introduce 50 mL of 50-mM aqueous tartaric acid solution. (Cleaning time: 165 minutes)
- (Attention)
- Keeping the Washing solution in the column for a long time will lead to column deterioration. Please replace the Washing solution with the eluent immediately after cleaning.