Please read the column cleaning method for your column.
Asahipak NH2Pseries
For an efficient cleaning, reverse the flow direction.
Flow rate during the cleaningScroll
Product name | NH2P-50 4B | NH2P-50 4D | NH2P-50 4E | NH2P-40 3E |
---|---|---|---|---|
Flow rate | 0.5 mL/min | 0.5 mL/min | 0.5 mL/min | 0.17 mL/min |
Product name | NH2P-50 2D | NH2P-40 2E | NH2P-50 10E | NH2P-90 20F |
---|---|---|---|---|
Flow rate | 0.1 mL/min | 0.1 mL/min | 1.5 mL/min | 5.0 mL/min |
Method 1: General cleaning Scroll
Washing solution | Cleaning time (min) | ||||
---|---|---|---|---|---|
NH2P-50 4B | NH2P-50 4D | NH2P-50 4E | NH2P-40 3E | ||
1 | H2O | 2 | 6 | 10 | 13 |
2 | 0.1 M NaOH aq. solution | 24 | 75 | 120 | 150 |
3 | H2O | 4 | 12 | 20 | 26 |
4 | Eluent | 12 | 40 | 60 | 75 |
Scroll
Washing solution | Cleaning time (min) | ||||
---|---|---|---|---|---|
NH2P-50 2D | NH2P-40 2E | NH2P-50 10E | NH2P-90 20F | ||
1 | H2O | 6 | 10 | 20 | 25 |
2 | 0.1 M NaOH aq. solution | 75 | 120 | 190 | 270 |
3 | H2O | 12 | 20 | 30 | 50 |
4 | Eluent | 40 | 60 | 100 | 140 |
Method 2: Adsorption of metal ions or in case Method 1 did not work Scroll
Washing solution | Cleaning time (min) | ||||
---|---|---|---|---|---|
NH2P-50 4B | NH2P-50 4D | NH2P-50 4E | NH2P-40 3E | ||
1 | H2O | 2 | 6 | 10 | 13 |
2 | 0.1 M HNO3 aq. solution | 24 | 75 | 120 | 150 |
3 | H2O | 2 | 6 | 10 | 13 |
4 | 0.1 M NaOH aq. solution | 24 | 75 | 120 | 150 |
5 | H2O | 4 | 12 | 20 | 26 |
6 | Eluent | 12 | 40 | 60 | 75 |
Scroll
Washing solution | Cleaning time (min) | ||||
---|---|---|---|---|---|
NH2P-50 2D | NH2P-40 2E | NH2P-50 10E | NH2P-90 20F | ||
1 | H2O | 6 | 10 | 20 | 25 |
2 | 0.1 M HNO3 aq. solution | 75 | 120 | 190 | 270 |
3 | H2O | 6 | 10 | 20 | 25 |
4 | 0.1 M NaOH aq. solution | 75 | 120 | 190 | 270 |
5 | H2O | 12 | 20 | 30 | 50 |
6 | Eluent | 40 | 60 | 100 | 140 |
- (Attention)
- Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.
- Strong alkaline solvents such as 0.1-M aqueous sodium hydroxide solution can damage the detector cell. Do not connect the detector while regenerating the column and collect elute directly from the column outlet to the waste container.