Comparison of Calibration Curves for KD-802.5 and SB-802.5 HQ (DMF Eluent: PEG, PEO and PS)

For the calibration curve for GPC KD-802.5, PEG and PEO should be used as the standards. The calibration curve with PS is far different from that with PEG and PEO. For the calibration curve for OHpak SB-802.5 HQ, both PEG, PEO and PS can be used as the standards. The calibataion curve with PS is almost as same as that with PEG and PEO.
Please refer to Polystyrene Standards (9) (KD-802).


Sample :
PEG, Poly(ethylene glycol)
PEO, Poly(ethylene oxide)
PS, Polystyrene


Columns      : Shodex GPC KD-802.5, OHpak SB-802.5 HQ (8.0 mm I.D. x 300mm each)
Eluent       : DMF
Column temp. : 40 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications

Applications (Related Information)