Analysis of Anions (SI-36 4D)

Please use this column with suppressor type ion chromatography system.

IC SI-36 4D is a column for anion analysis using potassium hydroxide aqueous solution as eluent. With columns using carbonate aqueous solution as eluent, sulfate ion elutes after phosphate ion. But using SI-36 4D, sulfate ion is eluted faster than monovalent bromide ions. In the case of columns using carbonate aqueous solution as eluent. Adding some organic solvent to the eluent is necessary to separating sulfite and sulfate ions. On the other hand, SI-36 4D does not require organic solvent to achive the separating. Under gradient conditions, phosphate ions are eluted within 15 minutes, analysis time is shortened.


Sample : 25 μL
1. F- 0.5 mg/L
2. Cl- 3 mg/L
3. NO2- 5 mg/L
4. SO32- 5 mg/L
5. SO42- 10 mg/L
6. Br- 10 mg/L
7. NO3- 10 mg/L
8. PO43- 20 mg/L


Column       : Shodex IC SI-36 4D (4.0 mm I.D. x 150 mm)
Eluent       : (Isocratic) 25 mM KOH aq. 
               (Gradient) KOH aq. 
                25 mM (0 to 6 min), 25 to 40 mM (6 to 8 min), 
                40 mM (8 to 15 min), 40 to 25 mM (15 to 15.01 min),
                25 mM (15.01 to 20 min
                (Eluent source : DionexTM EGC 500 KOH)
Flow rate    : 0.7 mL/min
Detector     : Suppressed conductivity
Column temp. : 30 °C