Rapid Analysis of Oxyhalides and Anions (SI-35 4D) (1)

Please use this column with suppressor type ion chromatography system.

The analytical performance of Shodex IC SI-35 4D (a column for rapid oxyhalides analysis) was compared to the existing oxyhalides-analysis column, SI-52 4E. The analysis time required by SI-35 4D is only one half of SI-52 4E 's analysis time.


Sample : (SI-35 4D) 20 μL, (SI-52 4E) 50 μL
1. F- 2 mg/L
2. ClO2- 1 mg/L
3. BrO3- 1 mg/L
4. Cl- 10 mg/L
5. NO2- 5 mg/L
6. Br- 10 mg/L
7. ClO3- 1 mg/L
8. NO3- 30 mg/L
9. HPO42- 15 mg/L
10. SO42- 40 mg/L

 

Column       : (Left) Shodex IC SI-35 4D (4.0 mm I.D. x 150 mm)
               (Right) Shodex IC SI-52 4E (4.0 mm I.D. x 250 mm)
Eluent       : 3.6 mM Na2CO3 aq.
Flow rate    : (Left) 0.6 mL/min
               (Right) 0.8 mL/min
Detector     : Suppressed conductivity
Column temp. : 45 °C