Anion Standards (20) (Simultaneous Analysis of HS-, SO32-, SO42- and S2O32-) (SI-90 4E)

Please use this column with suppressor type ion chromatography system.

Anion standards were analyzed using IC SI-90 4E (a column for anions analysis). HS- is difficult to detect by conductivity. To separate and detect the above-mentioned anions, UV and CD detector are connected in series.


Sample : 50 mg/L each, 20 μL
1. HS-
2. SO32-
3. SO42-
4. S2O32-


Column       : Shodex IC SI-90 4E (4.0 mm I.D. x 250 mm)
Eluent       : 1 mM Na2CO3 + 4 mM NaHCO3 + 5 % Acetone 
Flow rate    : 1.5 mL/min
Detector     : UV (210 nm) + Suppressed conductivity
Column temp. : 25 °C