Alkylamines were analyzed using IC YK-421 . It is possible to analyze alkylamines with short time and high sensitively, when acetonitrile was added to the eluent. Because hydrophobic cations (ex. Alkylamines) are eluted earlier with the addition of acetnitile to the eluent. Phosphate eluent is suitable for the analysis.
Sample : 50 μL
1. NH4+ 5 mg/L
2. Methylamine, MA 5 mg/L
3. Dimethylamine, DMA 5 mg/L
4. Trimethylamine, TMA 20 mg/L
Sample : 10 mg/L each, 50 μL
2. Methylamine, MA
5. Ethylamine, EA
6. Propylamine, PA
7. Butylamine, BA
Column : Shodex IC YK-421 (4.6 mm I.D. x 125 mm) Eluent : 4 mM H3PO4 aq./CH3CN=90/10 Flow rate : 1.0 mL/min Detector : Non-suppressed conductivity Column temp. : 25 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Monovalent and Divalent Cation Standards (1) (YK-421)
- Monovalent and Divalent Cation Standards (2) (YK-421)
- Monovalent and Divalent Cation Standards (3) (Dipicolinic Acid + Crown Ether Eluent) (YK-421)
- Monovalent and Divalent Cation Standards (4) (YK-421) (Phosphoric Acid Eluent)
- Monovalent and Divalent Cation Standards (5) (YK-421) (Phosphoric Acid + 18-Crown 6-Ether Eluent)
- Monovalent and Divalent Cation Standards (6) (Comparison of YS-50 and YK-421)
- Monovalent and Divalent Cation Standards (7) (Eluent Conteinig Acetonitrile) (YS-50)
- Monovalent and Divalent Cation Standards (8) (Ethylenediamine Containing Sample) (YS-50)
- Dynamic Range of Column (YK-421)
- High Sensitive Detection of Cations (YK-421)
- Effects of Acetonitrile Concentration on Elution Volume (YK-421)
- Effects of Methanesulfonic Acid Concentration on Elution Time (YS-50)
- Effects of Flow Rate on Plate Number (YS-50)
- Effects of Temperature (YS-50)
- Effects of Sample Load on Plate Number (YS-50)
- Linearity of Calibration Curves (YS-50)
- Effects of Crown Ethers (YS-50)
- Tap Water (1) (YK-421)
- Analysis of Tap Water (2) (YK-412) (Phosphoric Acid Eluent)
- Monovalent and Divalent Cations Analysis in Tap Water (Ethylenediamine Containing Sample) (YS-50)
- Monovalent and Divalent Cations Analysis in River Water (Ethylenediamine Containing Sample) (YS-50)
- Mineral Water (YS-50)
- Sea Water (2) (YS-50)
- Red Wine (1) (YK-421)
- Red Wine (3) (YS-50)
- Serum (YK-421)
- Long Chain Alkylamines (YK-421)
- Trialkylamines (YK-421)
- Tetraalkylammoniums (YK-421)
- Analysis of Tetraalkylammoniums (YS-50)
- Ethanolamines and Alkylamines (YK-421) (Phosphoric Acid + Acetonitrile)
- Analysis of Ethanolamines and Alkylamines (YS-50)
- Cations and Alkylamines (YS-50)
- Analysis of Methylamines (YS-50)
- Analysis of Ethylamines (YS-50)
- Analysis of Short Chain Alkylamines (YS-50)
- Cation Analysis in the Presence of PEG (YS-50)
- Cation Analysis in the Presence of Albumin (YS-50)
- Quantification of N-Methylpyrrolidine in Cefepime for Injection According to USP-NF Method (YK-421)