The National Ecological Environment Standards of China HJ 1362‑2024* “Determination of Phosphoric Acid Mist in Exhaust Gas from Stationary Sources by Ion Chromatography” specifies the use of a high‑capacity hydrophilic ion‑exchange column, which is packed with a base material such as polydivinylbenzene, ethylvinylbenzene, or polyvinyl alcohol, modified with quaternary ammonium groups having alkyl chains or alkyl alcohol chains. Using IC SI‑37 4D, an anion analysis column, a good linear calibration curve for phosphate and good separation of seven anions were obtained.
Note: The analytical conditions used in this application differ in part from those specified in the National Ecological Environment Standards of China HJ 1362‑2024.
*The version at the time of the application acquisition.
Sample: 25 µL
- 1.F- 1.0 mg/L
- 2.Cl- 3.0 mg/L
- 3.NO2- 3.0 mg/L
- 4.SO42- 5.0 mg/L
- 5.Br- 5.0 mg/L
- 6.NO3- 5.0 mg/L
- 7.PO43- 5.0 mg/L

Sample: 25 µL
- 7.PO43- 5.0 mg/L


- Column
- :Shodex IC SI-37 4D (4.0 mm I.D. x 150 mm)
- Eluent
- :23 mM KOH aq.
- Flow rate
- :0.7 mL/min
- Detector
- :Suppressed conductivity
- Column temp.
- :30 ℃