The National Ecological Environment Standards of China HJ 1361‑2024* “Determination of Nitric Acid Mist in Exhaust Gas from Stationary Sources by Ion Chromatography” specifies the use of a high‑capacity hydrophilic ion‑exchange column, which is packed with a base material such as polydivinylbenzene, ethylvinylbenzene, or polyvinyl alcohol, modified with quaternary ammonium groups having alkyl chains or alkyl alcohol chains. Using IC SI-90 4E, an anion analysis column, a good linear calibration curve for nitrate and good separation of seven anions were obtained.
In contrast to the official method, sulfate eluted last under the conditions used in this application.
*The version at the time of the application acquisition.
Sample: 25 µL
- 1.F- 1.0 mg/L
- 2.Cl- 3.0 mg/L
- 3.NO2- 3.0 mg/L
- 4.Br- 5.0 mg/L
- 5.NO3- 5.0 mg/L
- 6.HPO42- 5.0 mg/L
- 7.SO42- 5.0 mg/L

Sample: 25 µL
- 5.NO3- 5.0 mg/L


- Column
- :Shodex IC SI-90 4E (4.0 mm I.D. x 250 mm)
- Eluent
- :3.2 mM Na2CO3 + 1.0 mM NaHCO3 aq.
- Flow rate
- :0.7 mL/min
- Detector
- :Suppressed conductivity
- Column temp.
- :30 ℃