The National Ecological Environment Standards of China HJ 1288 2023* “Water Quality - Determination of Acrylic Acid by Ion Chromatography” specifies the use of an ion exchange column compatible with carbonate based eluents, which is packed with polymer-based packing material, such as styrene divinylbenzene or polyvinyl alcohol, modified with quaternary ammonium groups, or a column with equivalent performance. Using IC SI-90 4E, an anion analysis column, a good linear calibration curve for acrylic acid were obtained.
*The version at the time of the application acquisition.
Sample: 25 µL
- 1.Acrylic acid 5 mg/L


- Column
- :Shodex IC SI-90 4E (4.0mm I.D. x 250mm)
- Eluent
- :1.8 mM Na2CO3 + 1.7mM NaHCO3 aq.
- Flow rate
- :1.0 mL/min
- Detector
- :Suppressed conductivity
- Column temp.
- :30 ℃