Analysis of Mannitol in According to Chinese Pharmacopoeia 2025 Edition Standard (SC1011)

The Chinese Pharmacopoeia 2025 Edition*, Part IV, under “Mannitol” monograph, analysis of related substances by high performance liquid chromatography (HPLC) “Tests” section specifies the use of a column packed with strong cation exchange resin (calcium form) in which sulfonic acid groups are bonded to a cross linked styrene divinylbenzene copolymer or a column with equivalent separation performance. SUGAR SC1011 column confirmed the requirements were met. SC1011 column is also applicable to the “Assay” test.

System suitability requirementsResolution between mannitol and sorbitol: > 2.0
Resolution between maltitol and the first peak of isomalt: ≥ 1.5

*The version at the time of the application acquisition.

Sample: 25 µL
25 mg/mL each

  1. 1.Mannitol
  2. 2.Sorbitol
chromatogram of mannitol

Sample: 25 µL
1 mg/mL each

  1. 3.Isomalt (= Palatinit)
  2. 4.Maltitol
chromatogram of isomalt
Column
Shodex SUGAR SC1011 (8.0 mm I.D. x 300 mm)
Eluent
H2O
Flow rate
0.5 mL/min
Detector
RI (40 ℃)
Column temp.
80 ℃