• TOP
  • Applications
  • Analysis of Maltitol According to Chinese Pharmacopoeia 2025 Edition (SC1011)

Analysis of Maltitol According to Chinese Pharmacopoeia 2025 Edition (SC1011)

The Chinese Pharmacopoeia 2025 Edition*, Part IV, the monograph for Maltitol, analysis by high‑performance liquid chromatography specifies the use of a column packed with strong cation exchange resin (calcium form). SUGAR SC1011 column confirmed the requirement was met.

System suitability requirementResolution between maltitol and sorbitol: > 2

*The version at the time of the application acquisition.

Sample: 20 µL
5 mg/mL each

  1. 1.Maltitol
  2. 2.Sorbitol
Chromatogram of Maltitol
Column
Shodex SUGAR SC1011 (8.0 mm I.D. x 300 mm)
Eluent
H2O
Flow rate
0.5 mL/min
Detector
RI (35 ℃)
Column temp.
75 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications