• TOP
  • Applications
  • Analysis of Human Serum Albumin Multimers According to Chinese Pharmacopoeia 2025 Edition (KW-803)

Analysis of Human Serum Albumin Multimers According to Chinese Pharmacopoeia 2025 Edition (KW-803)

The Chinese Pharmacopoeia 2025 Edition*, Part IV, the high‑performance liquid chromatography method for “Human Serum Albumin Multimers” specifies the use of a high‑performance size‑exclusion chromatography (SEC) column packed with hydrophilic silica gel, with an exclusion limit of 300 kDa, a particle size of 10 µm, and dimensions of 7.5 mm (I.D.) x 60 cm (L)**. PROTEIN KW-803 column confirmed the requirements were met.

System suitability requirements

Resolution between the human serum albumin monomer peak and dimer peak: > 1.5
Tailing factor of the human serum albumin monomer peak: 0.95 - 1.40

*The version at the time of the application acquisition.
**Chinese Pharmacopoeia 2025 edition allows changes in column dimensions, packing material, and particle sizes. 

Sample: 20 µL
Human serum albumin 12 mg/mL

  1. 1.Dimer
  2. 2.Monomer
Column
Shodex PROTEIN KW-803 (8.0 mm I.D. x 300 mm)
Eluent
0.5 M NaH2PO4 aq./0.5 M Na2HPO4 aq./Isopropanol/H2O=200/420/15.5/914.5
Flow rate
0.6 mL/min
Detector
:UV (280 nm)
Column temp.
30 ℃