Instead of dichloromethane which may cause the environmental pollution, non-halogenated solvent such as ethyl acetate and acetone can be used as eluent. Each solvent is mixed with cyclohexane and used for CLNpak EV-2000 series as the eluent. The effect of the solvents and their mixing ratio is shown below. 
Sample :
1. 125 mg/mL Corn oil
2. 1 mg/2 mL β-Carotene
3. 1 mg/2 mL Fluvalinate
4. 0.3 mg/2 mL Chinomethionate
Columns : (Left); Shodex CLNpak EV-G AC (20 mm I.D. x 100 mm)
+ EV-2000 AC (20 mm I.D. x 300 mm)
(Center); Shodex CLNpak EV-2000 (20 mm I.D. x 300 mm)
(Right); Shodex CLNpak EV-G (20 mm I.D. x 100 mm)
+ EV-2000 (20 mm I.D. x 300 mm)
Eluent : (Left); Acetone/Cyclohexane
(Center); Ethyl acetate/Cyclohexane
(Right); Dichloromethane/Cyclohexane
Flow rate : 4.0 mL/min
Detector : UV (254 nm)
Column temp. : 25 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Sample Preparation for Animal and Fishery Products According to Japanese MHLW Notification
- GPC Clean-up for Animal and Fishery Products According to Japanese MHLW Notification (1) (EV-2000 AC)
- GPC Clean-up for Animal and Fishery Products According to Japanese MHLW Notification (2) (EV200AC-12F)
- Example of GPC Clean-up for Animal and Fishery Products According to Japanese MHLW Notification (1) (Medium Fatty Tuna and Beef Fat) (EV-2000 AC)
- Example of GPC Clean-up for Animal and Fishery Products According to Japanese MHLW Notification (2) (Milk, Egg) (EV-2000 AC)
- Elution Profile for Pesticides (EV-2000 AC)
- Effects of Temperature on GPC Clean-up
- Effects of Eluent Composition on GPC Clean-up
- Elution Volume for Lipid and Fatty Acids (EV-2000 AC)
- Separation with Acetone/Cyclohexane Eluent (EV-2000 AC)
- Separation with Ethyl Acetate/Cyclohexane Eluent (EV-2000)
- Spinach Pigment Removal (EV-2000 AC)
- Spinach Pigment Removal - Removal Rate (EV-2000 AC)
- Effects of Sample Injection Volume
- Effects of Flow Rate (EV-2000 AC)
- Effects of Column Temperature
- Fractionation Standards (EV-2000 AC)