Two different temperature settings (room temerature and 40 °C) were compared for the GPC clean-up method using CLNpakEV-2000 AC . At room temperature, the high viscosity causes peak broadening and increases the analysis time. Therefore, analysis at 40 °C is preferred. In Food Sanitation Law #1003001 (Released by Ministry of Health, labour and welfare in Japan, 2006), analysis temperature is set at 40 °C.
Sample : 5 mL
Fatty tuna
Acrinathrin 40 mg/L
Tricyclazole 40 mg/L
Columns : Shodex CLNpak EV-G AC (20.0 mm I.D. x 100mm) + EV-2000 AC (20.0 mm I.D. x 300 mm) Eluent : Acetone/Cyclohexane=1/4 Flow rate : 5 mL/min Detector : UV (254 nm) Column temp. : 40 °C or Rooom temp.
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Elution Profile for Pesticides (EV-2000 AC)
- Effects of Eluent Components
- Separation with Acetone/Cyclohexane Eluent (EV-2000 AC)
- Separation with Ethyl Acetate/Cyclohexane Eluent (EV-2000)
- Spinach Pigment Removal (EV-2000 AC)
- Spinach Pigment Removal - Removal Rate (EV-2000 AC)
- Effects of Sample Injection Volume
- Effects of Flow Rate (EV-2000 AC)
- Effects of Column Temperature
- Preparation of Test Samples for Agricultural Chemicals in Livestock and Marine Products
- GPC Clean-up of Pesticide Residues (EV-2000 AC)
- Fractionation Standards (EV-2000 AC)
- Elution Volume for Lipid and Fatty Acids (EV-2000 AC)
- Effects of Eluent Composition on GPC Clean-up
- GPC Clean-up Sample (1) (Fatty Tuna and Beef Tallow)
- GPC Clean-up Sample (2) (Milk and Egg)
- Preparation Range of Agricultural Chemicals (EV2000AC-12F)