Total cyanide in river water was analysed using DEAE-825 (a column for weak anion exchange chromatography) and an electrochemical detector (ECD).
Sample : distilled River water, 100 μL
Column : Shodex IEC DEAE-825 (8.0 mm I.D. x 75 mm) Eluent : (10 mM Sodium carbonate + 1 mM Ethylenediamine)/CH3OH=90/10 Flow rate : 1.0 mL/min Detector : ECD (Electrode; Ag, 0 mV SCE) Column temp. : 40 °C