Total Cyanide in River Water (DEAE-825)

Total cyanide in river water was analysed using DEAE-825 (a column for weak anion exchange chromatography) and an electrochemical detector (ECD).


Sample : distilled River water, 100 μL


 

Column       : Shodex IEC DEAE-825 (8.0 mm I.D. x 75 mm)
Eluent       : (10 mM Sodium carbonate + 1 mM Ethylenediamine)/CH3OH=90/10
Flow rate    : 1.0 mL/min
Detector     : ECD (Electrode; Ag, 0 mV SCE)
Column temp. : 40 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications