Amides (2) (Comparison of Reversed Phase Columns)

Using three reversed-phase columns, RSpak DM-614, RSpak DE-613, and RSpak DS-613, separation of alkylamides under identical conditions were evaluated. DM-614, which is packed with a highly polar packing material, shows weak retention, resulting in insufficient separation. In contrast, DS-613 provides strong retention, enabling separation, but with a longer analysis time. DE-613 exhibits moderate retention and achieves good separation. Under the given condition, DE-613 is considered the optimal column.

Sample: Amides

  1. 1.Acetamide
  2. 2.Propionamide
  3. 3.Butylamide
  4. 4.Valeramide
Column
:(A) Shodex RSpak DM-614, (B) Shodex RSpak DE-613, (C) Shodex RSpak DS-613 (6.0 mm I.D. x 150 mm each)
Eluent
:H2O/CH3OH=70/30
Flow rate
:1.0 mL/min
Detector
:UV (210 nm)
Column temp.
:40 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications