Analysis of Glyceryl Dibehenate According to USP-NF Method (KF-802)

According to the United States Pharmacopeia and the National Formulary (USP-NF 2026 Issue 5*), "ASSAY" for "Glyceryl Dibehenate" should be carried out with a 7.5 mm (I.D.) x 60 cm (length) column filled with L21 packing material. Use of two GPC KF-802 columns, 8.0 mm (I.D.) x 30 cm (length) each, filled with L21 packing material in series confirmed the requirement was met.

System suitability requirementThe retention times of the major peaks of the Sample solution correspond to those of the System suitability solution.
Note: The relative retention times for tri-, di-, monoglycerides, and glycerin are about 0.73, 0.76, 0.82, and 1.0,respectively.

*The version at the time of the application acquisition.

Sample: 40 µL
(Sample solution) 40 mg/mL Glyceryl Dibehenate in THF
(System suitability solution) 40 mg/mL USP Glyceryl Dibehenate in THF

  1. 1.Triglycerides
  2. 2.Diglycerides
  3. 3.Monoglycerides
  4. 4.Glycerin
chromatgram of glyceryl dibehenate
Column
:Shodex GPC KF-802 (8.0 mm I.D. x 300 mm) x 2
Eluent
:THF
Flow rate
:1.0 mL/min
Detector
:RI (35 ℃)
Column temp.
:35 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index