Analysis of Sucrose Palmitate According to USP-NF Method (KF-802)

According to the United States Pharmacopeia and the National Formulary (USPNF 2026 Issue 2*), determination of monoesters, diesters, triesters, and polyesters in the "ASSAY" for Sucrose Palmitate should be carried out with a column filled with L21 packing material (column size: 7 mm I.D. x 60 cm). Two GPC KF-802 columns (8.0 mm I.D. x 30 cm/column) filled with L21 packing material, connected in series achieved a good separation of diester, monoester, and palmitic acid.

*The version at the time of the application acquisition.

Sample: 20 µL
(Sample solution) 15 mg/mL of Sucrose palmitate in THF

  1. 1.Triesters and polyesters
  2. 2.Diesters
  3. 3.Monoester
  4. 4.Palmitic acid
Chromatogram of sucrose palmitate
Column
Shodex GPC KF-802 (8.0 mm I.D. x 300 mm) x 2
Eluent
THF
Flow rate
1.2 mL/min
Detector
RI
Column temp.
40 ℃

Sample Name Index