LC/MS Analysis of Carnitine and Trimethyl Glycine (NH2P-40 2D)

Carnitine and Trimethyl glycine were analyzed by Asahipak NH2P-40 2D a polymer-based amino column under HILIC mode. Hydrophilic compounds are retained at a high concentration of acetonitrile under HILIC mode conditions. This is advantageous for high sensitivity analysis by LC/MS.
*Sales of Asahipak NH2P-40 2D used in this application is discontinued. Please consider trying Asahipak NH2P-50 2D.

Sample: 5 μL
5 ng/mL each (in H2O/CH3CN=25/75)

  1. 1.Trimethyl glycine
  2. 2.Carnitine
Column
Shodex Asahipak NH2P-40 2D (2.0 mm I.D. x 150 mm)
Eluent
50 mM HCOONH4 aq./CH3CN=20/80
Flow rate
0.2 mL/min
Detector
ESI-MS (SIM Positive)
Column temp.
30 ℃