According to Japanese Testing Methods for Fertilizers (2016*), "High-performance liquid chromatography method" under "5.7 Sulfamic acid (amide sulfate)" should be carried out with a column filled with silica gel modified with crosslinked diol or polyhydroxy methacrylate. The method mentions that this analysis is applicable to fertilizers, and ODP2 HP-2D as a suitable column. In this application, sulfamic acid was analyzed using ODP2 HP-2D. The calibration curve for sulfamic acid showed high linearity with a coefficient of determination (R2 = 0.9977), over the concentration range 10 to 600 ng/mL. Although ODP2 HP is normally categorized as a reversed-phase column, because of its relatively polar packing material, it can also provide HILIC mode effects as shown in this data.
*The version at the time of the application acquisition.
Sample: 1 µL
- 1.Sulfamate 100 ng/mL


- Column
- :Shodex ODP2 HP-2D (2.0 mm I.D. x 150 mm)
- Eluent
- :100 mM Ammonium formate buffer (pH3.2)/CH3CN=10/90
- Flow rate
- :0.2 mL/min
- Detector
- :ESI-MS (SIM Negative)
- Column temp.
- :40 ℃
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- High Sensitive Analysis of Perchloric Acid and Oxyhalides by LC/MS (JJ-50 2D)
- LC/MS/MS Analysis of Bromate According to Japanese Water Supply Act (JJ-50 2D)
- LC/MS Analysis of Iodide (VT-50 2D)
- Simultaneous LC/MS Analysis of Haloacetic Acids and Anions Including Chlorate and Bromate (JJ-50 2D)
- LC/MS/MS Analysis of Chlorate and Bromate According to Japanese Water Supply Act (JJ-50 2D)
- Analysis of Perchloric Acid by LC/MS (SI-35 2B)