Anion Standards and Organic Acids (4) (SI-50 4D)

Please use this column with suppressor type ion chromatography system.

Anion standards and organic acids were analyzed using IC SI-50 4D (a column for anions analysis).

Sample: 20 µL

  1. 1.F- 2 mg/L
  2. 2.Acetate 10 mg/L
  3. 3.Formate 2 mg/L
  4. 4.Cl- 3 mg/L
  5. 5.NO2- 5 mg/L
  6. 6.Br- 10 mg/L
  7. 7.NO3- 10 mg/L
  8. 8.HPO42- 15 mg/L
  9. 9.SO42- 15 mg/L
Chromatgram of organic acids and anions
Column
Shodex IC SI-50 4D (4.0 mm I.D x 150 mm)
Eluent
3.2 mM Na2CO3 + 1.0 mM NaHCO3 aq.
Flow rate
0.7 mL/min
Detector
Supressed conductivity
Column temp.
30 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index