Effects of Column Temperature (SI-50 4D)

Please use this column with suppressor type ion chromatography system.

The effects of column temperature on analysis of seven general inorganic anions using IC SI-50 4D, a column for anions analysis, were studied. Effects of temperature on elution positions of each ion was found to be negligible. However, separation performance gradually decreased above 40 ℃. Therefore, it is recommended to set the column temperature between 30 and 35 ℃.

Sample: 20 μL

  1. 1.F- 2 mg/L
  2. 2.Cl- 3 mg/L
  3. 3.NO2- 5 mg/L
  4. 4.Br- 10 mg/L
  5. 5.NO3- 10 mg/L
  6. 6.HPO42- 15 mg/L
  7. 7.SO42- 15 mg/L
chromatogram of SI-50 4D

 

chromatogram of SI-50 4D
Column
Shodex IC SI-50 4D (4.0 mm I.D. x 150 mm)
Eluent
:3.2 mM Na2CO3 + 1.0 mM NaHCO3 aq.
Flow rate
0.7 mL/min
Detector
Suppressed conductivity
Column temp.
30, 35, 40, 45, 50 ℃