Please use this column with suppressor type ion chromatography system.
Recently ozonation has been applied for sterilization of drinking water. Ozonation transforms part of bromide to bromate which is a potential carcinogen. So, it is very important to measure bromate in drinking water. A modeldrinking water with bromate was measured by suppressed coductivity with IC SI-50 4E.
Sample : Model drinking water, 500 μL
1. 0.1 mg/L F-
2. 0.003 mg/L BrO3-
3. 20 mg/L Cl-
4. 0.1 mg/L Br-
5. 10 mg/L NO3-
6. 10 mg/L SO42-
1. : 0.02 mg/L
2. : 0.01 mg/L
3. : 0.003 mg/L
Column : Shodex IC SI-50 4E (4.0 mm I.D. x 250 mm) Eluent : 3.2 mM Na2CO3 + 1 mM NaHCO3 aq. Flow rate : 0.8 mL/min Detector : Suppressed conductivity Column temp. : 25 °C