Anions in milk were analyzed using IC I-524A (an anion analysis column for non-suppressor method). Milk contains a considerable amount of proteins and fats, and therefore should be pretreated. 100-fold dilution with pure water was made and then treated by an solid phase extraction cartridge for reversed phase chromatography.
Sample : 100-fold diluted Milk, 50 μL
2. HCO3-
3. Cl-
6. NO3-
7. HPO42-
Sample : Anion Standards, 50 μL
1. F- 5 mg/L
2. HCO3- 30 mg/L
3. Cl- 5 mg/L
4. NO2- 5 mg/L
5. Br- 10 mg/L
6. NO3- 10 mg/L
7. HPO42- 10 mg/L
8. SO42- 10 mg/L
Column : Shodex IC I-524A (4.6 mm I.D. x 100 mm) Eluent : 1.5 mM p-Hydoxybenzoic acid + 1.7 mM N,N-Diethylethanolamine + 10 % CH3OH Flow rate : 1.5 mL/min Detector : Non-suppressed coductivity Column temp. : 40 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Anion Standards (1) (I-524A)
- Anion Standards (2) (I-524A)
- Anion Standards (3) (I-524A)
- Anion Standards (6) (I-524A)
- Anion Standards (7) (Anions and Organic Acids) (I-524A)
- Anion Standards (8) (Eluent : Octanesulfonic Acid and Boric Acid) (I-524A)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Polyphosphoric Acids (I-524A)
- Effects of pH on Retention Time (1) (I-524A)
- Dynamic Range of Column (I-524A)
- Effects of Sample Load on Column Efficiency (1) (I-524A)
- Tap Water (3) (I-524A)
- Tap Water (4) (I-524A)
- River Water (1) (I-524A)
- High Sensitive Detection Method for Anions (I-524A)
- Carbonate and Other Anions (1) (I-524A)
- Anions in Beer (I-524A)
- Chinese Cabbage (I-524A)
- Iodic Ion in Urine (I-524A)
- Sea Water (1) (I-524A)
- Sodium Azide (1) (I-524A)
- Lithium Tetrafluoroborate (I-524A)