Anions of chinese cabbage are analyzed using IC I-524A (an anion analysis column for non-suppressor method). Anions in the sample were extracted using a juicer-mixer and then filtered.
(Preparation of eluent)
(1) Put 0.415 g of reagent grade phthalic acid and 0.278 g of reagent grade tris(hydroxymethyl)aminomethane into a 1liter volumetric flask.
(2) Dilute with purified water to make up a 1 liter solution.
(3) Dissolve using an ultrasonic vibrator.
(4) Pass through a 0.2 μm membrane filter before use.
Sample : Chinese cabbage
(A); Leaves, (B); Stalks
1. H2PO4-
2. Cl-
3. NO3-
4. SO42-
Column : Shodex IC I-524A (4.6 mm I.D. x 100 mm) Eluent : 2.5 mM Phthalic acid (pH4.0 adjusted by Tris(hydroxymethyl)aminomethane) Flow rate : 1.2 mL/min Detector : Non-suppressed coductivity
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Anion Standards (1) (I-524A)
- Anion Standards (2) (I-524A)
- Anion Standards (3) (I-524A)
- Anion Standards (6) (I-524A)
- Anion Standards (7) (Anions and Organic Acids) (I-524A)
- Anion Standards (8) (Eluent : Octanesulfonic Acid and Boric Acid) (I-524A)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Polyphosphoric Acids (I-524A)
- Effects of pH on Retention Time (1) (I-524A)
- Dynamic Range of Column (I-524A)
- Effects of Sample Load on Column Efficiency (1) (I-524A)
- Tap Water (3) (I-524A)
- Tap Water (4) (I-524A)
- River Water (1) (I-524A)
- High Sensitive Detection Method for Anions (I-524A)
- Carbonate and Other Anions (1) (I-524A)
- Anions in Beer (I-524A)
- Iodic Ion in Urine (I-524A)
- Milk (I-524A)
- Sea Water (1) (I-524A)
- Sodium Azide (1) (I-524A)
- Lithium Tetrafluoroborate (I-524A)
- Analysis of Hydrophobic Anions (I-524A)