Tap Water (3) (I-524A)

Anions in tap water were analyzed using IC I-524A (an anion analysis column for non-suppressor method). Since phosphoric ion is not present in this sample, F-, Cl-, NO3- and SO42- can be separated in a short time without t he interference of phosphoric ion peak using this eluent. When both F- and H2PO4- are present, they cannot be separated each othe with this analysis conditionr.


Sample : Tap water, 100 μL
1. F-
2. Cl-
3. NO3-
4. SO42-


Column       : Shodex IC I-524A (4.6 mm I.D. x 100 mm)
Eluent       : 2.5 mM Phthalic acid + 2.3 mM Tris(hydroxymethyl)aminomethane aq.
Flow rate    : 1.2 mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications