The eluent used here is suitable for the simultaneous analysis of inorganic anions and dibasic and tribasic organic acids using IC I-524A (an anion analysis column for non-suppressor method). Organic acids are contained in beer and they interfere with the quantification of anions. Cl-, NO3- and SO42- can be separated without such interference of organic acids. Organic acids, oxalic acid and citric acid also can be analyzed without interference.
Sample : 10-fold diluted Beer , 100 μL
1. Acetic acid
2. H2PO4- , Succinic acid , Pyroglutamic acid
3. Lactic acid , Pyruvic acid
4. Cl-
5. Malic acid , Br-
6. NO3-
7. Oxalic acid
8. Citric acid
9. SO42-
Column : Shodex IC I-524A (4.6 mm I.D. x 100 mm) Eluent : 1.5 mM Phthalic acid + 1.38 mM Tris(hydroxymethyl)aminomethane + 300 mM Boric acid aq. Flow rate : 1.2 mL/min Detector : Non-suppressed coductivity Column temp. : 40 °C
Sample Name Index
- Acetic acid
- Beer
- Bromide, Br-
- Chloride, Cl-
- Citric acid
- Lactic acid
- Malic acid
- Nitrate, NO3-
- Oxalic acid
- Phosphate monobasic, H2PO4-
- Pyroglutamic acid
- Pyruvic acid
- Succinic acid
- Sulfate, SO42-
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Anion Standards (1) (I-524A)
- Anion Standards (2) (I-524A)
- Anion Standards (3) (I-524A)
- Anion Standards (6) (I-524A)
- Anion Standards (7) (Anions and Organic Acids) (I-524A)
- Anion Standards (8) (Eluent : Octanesulfonic Acid and Boric Acid) (I-524A)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Polyphosphoric Acids (I-524A)
- Effects of pH on Retention Time (1) (I-524A)
- Dynamic Range of Column (I-524A)
- Effects of Sample Load on Column Efficiency (1) (I-524A)
- Tap Water (3) (I-524A)
- Tap Water (4) (I-524A)
- River Water (1) (I-524A)
- High Sensitive Detection Method for Anions (I-524A)
- Carbonate and Other Anions (1) (I-524A)
- Chinese Cabbage (I-524A)
- Iodic Ion in Urine (I-524A)
- Milk (I-524A)
- Sea Water (1) (I-524A)
- Sodium Azide (1) (I-524A)
- Lithium Tetrafluoroborate (I-524A)
- Analysis of Hydrophobic Anions (I-524A)