Based on the method described in the Japanese regulation for water supply, anions were analyzed using IC I-524A (an anion analysis column for non-suppressor method).
Please refer to 'Regulation for Water Supply '.
Sample : 100 μL each
Anion standards solution
1. H2PO4- 10 mg/L
2. F- 1 mg/L
3. Cl- 1 mg/L
4. NO2- 5 mg/L
5. Br- 5 mg/L
6. NO3- 5 mg/L
7. SO42- 5 mg/L
Tap water
Column : Shodex IC I-524A (4.6 mm I.D. x 100 mm) Eluent : 2.0 mM Phthalic acid + 1.84 mM Tris(hydroxymethyl)aminomethane + 300 mM Boric acid aq. Flow rate : 1.2 mL/min Detector : Non-suppressed coductivity Column temp. : 40 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Anion Standards (1) (I-524A)
- Anion Standards (2) (I-524A)
- Anion Standards (3) (I-524A)
- Anion Standards (6) (I-524A)
- Anion Standards (7) (Anions and Organic Acids) (I-524A)
- Anion Standards (8) (Eluent : Octanesulfonic Acid and Boric Acid) (I-524A)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Polyphosphoric Acids (I-524A)
- Effects of pH on Retention Time (1) (I-524A)
- Dynamic Range of Column (I-524A)
- Effects of Sample Load on Column Efficiency (1) (I-524A)
- Tap Water (3) (I-524A)
- River Water (1) (I-524A)
- High Sensitive Detection Method for Anions (I-524A)
- Carbonate and Other Anions (1) (I-524A)
- Anions in Beer (I-524A)
- Chinese Cabbage (I-524A)
- Iodic Ion in Urine (I-524A)
- Milk (I-524A)
- Sea Water (1) (I-524A)
- Sodium Azide (1) (I-524A)
- Lithium Tetrafluoroborate (I-524A)
- Analysis of Hydrophobic Anions (I-524A)