Effects of pH on Retention Time (1) (I-524A)

Anion standards were analyzed using IC I-524A, an anion analysis column for non-suppressor method. This figure shows the effect of pH on retention time with a 2.5 mM phthalic acid eluent. Generally, the retention time decreases with the increase of pH. A peak for MoO42- cannot be obtained below pH 5.5, neither a peak for CrO42- below pH 6.0, nor a peak for H2PO4- can be obtained above pH 4.5.

Sample: 20 μL

  1. ClO4- 50 mg/L
  2. SCN- 50 mg/L
  3. I- 20 mg/L
  4. CrO42- 50 mg/L
  5. MoO42- 50 mg/L
  6. S2O32- 50 mg/L
  7. SO42- 10 mg/L
  8. NO3- 10 mg/L
  9. Br- 10 mg/L
  10. NO2- 10 mg/L
  11. Cl- 10 mg/L
  12. F- 10 mg/L
  13. H2PO4- 10 mg/L
Column
Shodex IC I-524A (4.6 mm I.D. x 100 mm)
Eluent
2.5 mM Phthalic acid (pH variable)
Flow rate
1.2 mL/min
Detector
Non-suppressed conductivity
Column temp.
40 ℃