Anion standards were analyzed using IC I-524A, an anion analysis column for non-suppressor method. This figure shows the effect of pH on retention time with a 2.5 mM phthalic acid eluent. Generally, the retention time decreases with the increase of pH. A peak for MoO42- cannot be obtained below pH 5.5, neither a peak for CrO42- below pH 6.0, nor a peak for H2PO4- can be obtained above pH 4.5.
Sample: 20 μL
- ClO4- 50 mg/L
- SCN- 50 mg/L
- I- 20 mg/L
- CrO42- 50 mg/L
- MoO42- 50 mg/L
- S2O32- 50 mg/L
- SO42- 10 mg/L
- NO3- 10 mg/L
- Br- 10 mg/L
- NO2- 10 mg/L
- Cl- 10 mg/L
- F- 10 mg/L
- H2PO4- 10 mg/L

- Column
- :Shodex IC I-524A (4.6 mm I.D. x 100 mm)
- Eluent
- :2.5 mM Phthalic acid (pH variable)
- Flow rate
- :1.2 mL/min
- Detector
- :Non-suppressed conductivity
- Column temp.
- :40 ℃
Sample Name Index
- Bromide, Br-
- Chloride, Cl-
- Chromate, CrO42-
- Fluoride, F-
- Iodide, I-
- Molybdate, MoO42-
- Nitrate, NO3-
- Nitrite, NO2-
- Perchlorate, ClO4-
- Phosphate monobasic, H2PO4-
- Sulfate, SO42-
- Thiocyanate, SCN-
- Thiosulfate, S2O32-
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Anion Standards (1) (I-524A)
- Anion Standards (2) (I-524A)
- Anion Standards (3) (I-524A)
- Anion Standards (6) (I-524A)
- Anion Standards (7) (Anions and Organic Acids) (I-524A)
- Anion Standards (8) (Eluent : Octanesulfonic Acid and Boric Acid) (I-524A)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Polyphosphoric Acids (I-524A)
- Dynamic Range of Column (I-524A)
- Effects of Sample Load on Column Efficiency (1) (I-524A)
- Tap Water (3) (I-524A)
- Tap Water (4) (I-524A)
- River Water (1) (I-524A)
- High Sensitive Detection Method for Anions (I-524A)
- Carbonate and Other Anions (1) (I-524A)
- Anions in Beer (I-524A)
- Chinese Cabbage (I-524A)
- Iodic Ion in Urine (I-524A)
- Milk (I-524A)
- Sea Water (1) (I-524A)
- Sodium Azide (1) (I-524A)
- Lithium Tetrafluoroborate (I-524A)
- Analysis of Hydrophobic Anions (I-524A)