Anion standards were analyzed using IC I-524A. This eluent is suitable for the separation of ions that elute before chloride. After each sample is analyzed, it is necessary to wash the column with 200 to 300 μL of 0.1 M tartaric acid when the sample contains NO3-, SO42-, etc. because they elute very slowly and may be remaining inside the column.
Sample : 10 mg/L each, 50 μL
1. Acetic acid
2. H2PO4-
3. H2PO2-
4. H2PO3-
5. ClO2-
6. BrO3-
7. Cl-
Column : Shodex IC I-524A (4.6 mm I.D. x 100 mm) Eluent : 1 mM Tartaric acid aq. Flow rate : 1.2 mL/min Detector : Non-suppressed coductivity Column temp. : 40 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Anion Standards (1) (I-524A)
- Anion Standards (2) (I-524A)
- Anion Standards (6) (I-524A)
- Anion Standards (7) (Anions and Organic Acids) (I-524A)
- Anion Standards (8) (Eluent : Octanesulfonic Acid and Boric Acid) (I-524A)
- Anion Standards (15) (Comparison of NI-424 and I-524A)
- Polyphosphoric Acids (I-524A)
- Effects of pH on Retention Time (1) (I-524A)
- Dynamic Range of Column (I-524A)
- Effects of Sample Load on Column Efficiency (1) (I-524A)
- Tap Water (3) (I-524A)
- Tap Water (4) (I-524A)
- River Water (1) (I-524A)
- High Sensitive Detection Method for Anions (I-524A)
- Carbonate and Other Anions (1) (I-524A)
- Anions in Beer (I-524A)
- Chinese Cabbage (I-524A)
- Iodic Ion in Urine (I-524A)
- Milk (I-524A)
- Sea Water (1) (I-524A)
- Sodium Azide (1) (I-524A)
- Lithium Tetrafluoroborate (I-524A)
- Analysis of Hydrophobic Anions (I-524A)