Analysis of IRGANOX® 1010 in Cup of Instant Noodles (Styrene Foam) (GPC/MS) (KF-402HQ)

IRGANOX® 1010 in styrene foam cup for instant noodles was analyzed with GPC/MS using KF-402HQ (a semi-micro organic SEC (GPC) column). Only a low molecule weight domain can be poured into a mass spectrometer (MS) by using GPC mode. KF-402HQ can provide the high sensitivity analysis without pretreatment of samples.
Please refer to Simultaneous Analysis of Various Additives of Polymer in Polystyrene (GPC/MS).


Sample : 0.1 %, 5 μL
1. Polystyrene
2. IRGANOX® 1010

(HPLC)
Instrument   : Agilent 1100
Column       : Shodex GPC KF-402HQ (4.6 mm I.D. x 250 mm) x 2
Eluent       : THF
Flow rate    : 0.3 mL/min
Detector     : UV (210 nm)
Column temp. : 40 °C
 
(MS)
Instrument      : Thermo Quest LCQ Advantage
Ionization      : APCI
Capillary temp. : 180 °C
Vaporizer temp. : 450 °C
Source voltage  : 6000 V
Sheath gas flow : 80 L/min