Analysis of Antioxidant in Polystyrene (GPC/MS) (KF-402HQ)

Antioxidant (IRGANOX ® 1010) in Polystyrene was analyzed with GPC/MS using GPC KF-402HQ (a semi-micro organic SEC (GPC) column). Only a low molecule weight domain can be poured into a mass spectrometer (MS) by using GPC mode. Therefore, it is not necessary to pretreat.
Please refer to Simultaneous Analysis of Various Additives of Polymer in Polystyrene (GPC/MS)


Sample : 1 ppm IRGANOX ® 1010 in 1000ppm Polystyrene, 5 μL
1. Polystyren
2. IRGANOX ® 1010

(HPLC)
Instrument   : Agilent 1100
Column       : Shodex GPC KF-402HQ (4.6 mm I.D. x 250 mm) x 2
Eluent       : THF
Flow rate    : 0.3 mL/min
Detector     : UV (210 nm)
Column temp. : 40 °C
(MS) Instrument : Thermo Quest LCQ Advantage Ionization : APCI Capillary temp. : 180 °C Vaporizer temp. : 450 °C Source voltage : 6000 V Sheath gas flow : 80 L/min