Comparison of Calibration Curves for KD-800 Series and SB-800 HQ Series (DMF Eluent)

Print

Calibration curves for GPC KD-800 series and OHpak SB-800 HQ series are shown. Compared with the KD-800 columns, the SB-800 HQ columns perform better over a low molecular weight range.


Sample :
PEO, Poly(ethylene oxide)
PEG, Poly(ethylene glycol)


Columns      : Shodex OHpak SB-805 HQ + SB-802.5 HQ (8.0mmI.D. x 300mm each)
               Shodex GPC KD-805 + KD-802.5 (8.0mmI.D. x 300mm each)
Eluent       : DMF
Column temp. : 40°C