Vinylidene Chloride Acrylonitrile Copolymer (2) (Comparison of KD-800 and SB-800 HQ)

Vinylidene chloride acrylonitrile copolymer was analyzed. Compared with the GPC KD-800 columns, the OHpak SB-800 HQ columns perform better over a molecular weight range.


Sample : Vinylidene chloride acrylonitrile copolymer


Columns      : Shodex GPC KD-805 + KD-802.5, OHpak SB-805 HQ + SB-802.5 HQ
               (8.0 mm I.D. x 300 mm each)
Eluent       : 5 mM LiBr in DMF
Flow rate    : 1.0 mL/min
Detector     : RI
Column temp. : 40 °C