Polystyrene Oligomers' Separation Performance Comparison between FP-2002 and Other Manufacturer's Column

The polystyrene oligomers’ separation performance of GPC FP-2002, a column for organic SEC (GPC), is compared with that of other manufacturer’s rapid separation columns  (molecular weight exclusion limits (ML):5,000 and 1,000). Even FP-2002 has higher exclusion limit than the competitor's column, it has superior separation in the low molecular weight range.


Sample : 1 mL
Polystyrene (Mp: 580) 1.0 %
7mer (MW: 786)
6mer (MW: 682)
5mer (MW: 578)
4mer (MW: 474)
3mer (MW: 370)
2mer (MW: 266)


Columns      : Shodex GPC FP-2002, Column (A), (B) from other manufacturer 
               (20.0 mm I.D. x 600 mm each)
Eluent       : CHCl3
Flow rate    : 10.0 mL/min
Detector     : UV (254 nm)
Column temp. : 30 °C