The separation performance of GPC FP-2002, a column for organic SEC (GPC) is compared with that of other manufacturer’s rapid separation columns (molecular weight exclusion limits (ML): 5,000 and 1,000). FP-2002 has a larger pore volume than other manufacturer's columns and covers a wide molecular weight range.
Sample : 1 mL each
1. Polystyrene (Mp: 18,000) 0.5 %
2. Polystyrene (Mp: 9,820) 0.5 %
3. Polystyrene (Mp: 2,790) 1.0 %
4. Polystyrene (Mp: 1,300) 1.0 %
5. Polystyrene (Mp: 860) 1.0 %
6. Polystyrene (Mp: 580) 1.0 %
7. Benzene (MW: 78) 1.0 %
Columns : Shodex GPC FP-2002, Column (A), (B) from other manufacturer (20.0 mm I.D. x 600 mm each) Eluent : CHCl3 Flow rate : 10.0 mL/min Detector : UV (254 nm) Column temp. : 30 °C
Sample Name Index
Operation Manual / Certificate of Analysis
Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.
Product Name Index
Applications
- Polystyrene Standards (8) (K-804L)
- Polystyrene Standards (17) (K-806M)
- Polystyrene Standards (21) (GF-HQ : THF, Chloroform and DMF)
- Epoxy Resin (2) (K-803L)
- Poly(o-Cresyl Glycidyl Ether)-co-Formaldehyde (GF-310 HQ)
- Poly(Ether Ether Ketone) (K-806M)
- Poly(Butylene Terephthalate) (2) (Eluent: Chloroform) (K-806M)
- Low MW Polydimethylsiloxane (Chloroform) (K-801)
- Liquid Paraffin (KF-401HQ)
- Biodegradable Polymer (2) (Polylactic Acid) (LF-604)
- Biodegradable Polymer (3) (Polylactic Acid) (K-805L)
- Biodegradable Polymer (6) (Polylactic Acid) (KF-806M)
- Recycle Analysis of Lauryl Stearyl Thiodipropionate (K-2001)
- Comparison of Recycling Separation (FP-2002)
- Effects of Flow Rate (1) (FP-2002)
- Effects of Flow Rate (2) (FP-2002)
- Polystyrene Oligomers' Separation Performance Comparison between FP-2002 and Other Manufacturer's Column
- Analysis of Polyhydroxybutyric Acid (LF-804)