Effects of Sample Concentration (KF-606M)

Print

For GPC KF-606M, the effect of sample concentration on lution volume and peak-width is shown here. The effect of sample concentration is marked for the sample of larger molecular weight. When the injection volume is 10 μL, the sample concentration smaller than 0.05% is recommended for the sample of which maolecular weight is smaller than 6,750,000.

Sample : PS, Polystyrene, 10μL



Column       : Shodex GPC KF-606M (6.0mmI.D. x 150mm)
Eluent       : THF 
Column temp. : 40°C

Contact us