Effects of Solvent-Peak Separation Column on Calibration Curve

Around the range of molecular weight 100, a solvent peak and/or other peaks coming from polymer additives may appear. For improving the separation in the low molecular weight range without affecting the separation in the high molecular weight range, it is recommended to use a solvent-peak separation column, GPC KF-800D. The solvent-peak separation column should be used with linear type columns.


Sample : PS, Polystyrene

Columns      : (1); Shodex GPC KF-806L (8.0 mm I.D. x 300 mm)
               (2); Shodex GPC KF-806L (8.0 mm I.D. x 300 mm) + KF-800D (8.0 mm I.D. x 100 mm)
Eluent       : THF
Flow rate    : 1.0 mL/min
Detector     : RI