Ascorbic acid and isoascorbic acid were separatd using two columns of RSpak DE-413 (a column for polymer-based reversed phase chromatography). The calibration curve of ascorbic acid and isoascorbic acid shows good linearity respectively in the range from 1 to 100 μg/mL.
Sample : 10 μL
5 μg/mL each (in 0.1 % metaphosphoric acid)
1. Ascorbic acid
2. Isoascorbic acid
Column : Shodex RSpak DE-413 (4.6 mm I.D. x 150 mm) x 2 Eluent : 10 mM H3PO4 aq. Flow rate : 0.6 mL/min Detector : UV (254 nm) Column temp. : 40 ℃
Please select your location to continue:
For all other locations please click “close” and return to the homepage.
This website stores cookies on your computer. These cookies are used to collect information about how you interact with our website and allow us to remember you. We use this information in order to improve our services. If you are happy to accept these cookies please click “Accept” or simply continue to use our website. To find out more about the cookies we use, see “More Information”.