LC/MS Analysis of Choline and Acetylcholine (VC-50 2D)

Choline and acetylcholine were analyzed by LC/MS method using HILICpak VC-50 2D, a polymer-based HILIC column. The calibration curves of choline and acetylcholine show good linearity respectively in the range from 1 to 100 mM.


Sample : 5 μL
5 nM each (in H2O)
1. Acetylcholine
2. Choline



Column       : Shodex HILICpak VC-50 2D (2.0 mm I.D. x 150 mm)
Eluent       : 200 mM HCOOH aq./CH3CN=40/60
Flow rate    : 0.3 mL/min
Detector     : ESI-MS/MS (MRM Positive)
Column temp. : 40 °C