Choline and acetylcholine were analyzed by LC/MS method using HILICpak VC-50 2D, a polymer-based HILIC column. The calibration curves of choline and acetylcholine show good linearity respectively in the range from 1 to 100 mM.
Sample: 5 µL
5 nM each (in H2O)
- 1.Acetylcholine
- 2.Choline


- Column
- :Shodex HILICpak VC-50 2D (2.0 mm I.D. x 150 mm)
- Eluent
- :200 mM HCOOH aq./CH3CN=40/60
- Flow rate
- :0.3 mL/min
- Detector
- :ESI-MS/MS (MRM Positive)
- Column temp.
- :40 ℃