Choline and acetylcholine were analyzed using IC YS-50, a cation analysis column, under cation exchange mode. If high sensitive analysis is required, LC/MS/MS analysis using HILICpak VC-50 2D, a polymer-based HILIC mode column, is recommended. When VC-50 2D is used, the elution order of a sample is reversed when compared to HILIC mode.
Refer to: LC/MS Analysis of Choline and Acetylcholine (VC-50 2D)
Sample: 20 µL
10 µg/mL each (in H2O)
- 1.Choline
- 2.Acetylcholine

- Column
- :Shodex IC YS-50 (4.6 mm I.D. x 125 mm)
- Eluent
- :10 mM H3PO4 aq.
- Flow rate
- :1.0 mL/min
- Detector
- :Non-suppressed conductivity, UV (210 nm)
- Column temp.
- :40 ℃