Analysis of Choline and Acetylcholine (YS-50)

​Choline and acetylcholine were analyzed using IC YS-50, a cation analysis column, under cation exchange mode. If high sensitive analysis is required, LC/MS/MS analysis using HILICpak VC-50 2D, a polymer-based HILIC mode column, is recommended. When VC-50 2D is used, the elution order of a sample is reversed when compared to HILIC mode.
Refer to: LC/MS Analysis of Choline and Acetylcholine (VC-50 2D)

Sample: 20 µL
10 µg/mL each (in H2O)

  1. 1.Choline
  2. 2.Acetylcholine
Column
Shodex IC YS-50 (4.6 mm I.D. x 125 mm)
Eluent
10 mM H3PO4 aq.
Flow rate
1.0 mL/min
Detector
Non-suppressed conductivity, UV (210 nm)
Column temp.
40 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications

Applications (Related Information)