Human Mercaptoalbumin and Human Nonmercaptoalbumin in Human Serum

The ratio of human mercaptoalbumin(HMA) to human nonmercaptoalbumin(HNA) in serum is thought to have some relation with aging. The correlation between the ratio and various types of illness is being studied. Ion exchange mode has been used most commonly for the analysis of albumin, however, as isocratic elution widens the peaks, gradient elution must be applied in ion exchange mode. In contrast, one of the characteristics of Asahipak GS-520 ( a multimode column ) is its capacity of enabling isocratic elution of HMA and HNA with isocratic elution using simple eluent such as phosphate buffer. This is the original Asahipak GS-520 column separation method which allows proteins other than albumin to elute first in the GFC mode and only albumin to elute later.
GS-520 7E is an old type of GS-520 HQ.


Sample : Human serum
HMA (Human Mercaptoalbumin)
HNA (Human Nonmercaptoalbumin)
fHMA = (HMA)/(HMA + HNA)


Columns      : Shodex Asahipak GS-520 7E (7.5 mm I.D. x 250 mm) x 8
Eluent       : 0.03 M Sodium phosphate buffer (pH6.87) + 0.15 M Na2SO4
Flow rate    : 0.8 mL/min
Detector     : UV (280 nm)
Column temp. : 28 °C

Coutesy of Dr. Sogami, Fujita Health University, School of health Science