Five kinds of standard proteins were analyzed using IEC SP-FT 4A, a non-porous type column for cation exchange chromatography. These proteins can be separated well at 3.0 mL/min flow rate while maintaining the same separation at 1.5 mL/min flow rate. This analysis can be completed, within a mere 50 seconds.
Sample: 5 μL
-
- 1.Ovalbumin 0.8 mg/mL
- 2.Trypsinogen 1.6 mg/mL
- 3.Ribonuclease A 1.6 mg/mL
- 4.Cytochrome c 0.8 mg/mL
- 5.Lysozyme 0.4 mg/mL

- Column
- :Shodex IEC SP-FT 4A (4.6 mm I.D. x 10 mm ; 2.7 μm)
- Eluent
- :(A); 20 mM *MES buffer (pH6.0)
(B); (A) + 0.5 M Na2SO4 *MES : 2-(N-Morpholino)ethanesulfonic acid
Linear gradient; (A) to (B) (2 min) - Flow rate
- :0.5 mL/min, 3.0 mL/min
- Detector
- :UV (280 nm)
- Column temp.
- :Room temp.
Sample Name Index
Operation Manual / Certificate of Analysis